• Fluke 725Ex Intrinsically Safe Multifunction Process Calibrator

  • Đăng ngày 11-07-2017 04:48:52 PM - 62 Lượt xem
  • Giá bán: Liên hệ

  • The Simply Powerful Intrinsic Safe Calibration Tool

     

    The 725Ex Intrinsically Safe Multifunction Process Calibrator is powerful yet easy-to-use. Combined with one of the (8) Fluke-700PEx pressure modules, the 725Ex is able to calibrate almost any process instrument likely to need service in an area where explosive gasses may be present.

    Measure and source mA, volts, temperature (RTDs and thermocouples), frequency, ohms, and pressure, using optional pressure modules.

    Do you need to calibrate transmitters? The 725Ex split display lets you view input and output values simultaneously. For valve and I/P tests, you can source mA while measuring pressure. The 725Ex has auto-stepping and auto-ramping for remote testing, plus 25% stepping for fast linearity tests.

    From the moment you pick it up, the Fluke 725Ex is ready to perform. Its simple controls, without menus, make operation easy. Its memory functions make set up fast. And its rugged design lets it work as hard as you do.

        


Số lượng

Use the measure and source functions to test and calibrate almost any process parameter. The Fluke 725Ex Intrinsically Safe Calibrator is a powerful, multifunction calibration solution that offers:

 

  • ATEX II 1 G Ex ia IIB 171°C KEMA 04ATEX 1303X
  • I.S. Class I, Division 1 Groups B-D, 171°C compliance
  • Measure Volts dc, mA, RTDs, thermocouples, frequency and ohms
  • Source or simulate Volts dc, mA, RTDs, thermocouples, frequency and ohms
  • Two channel simultaneous source and measure capability for calibration of transmitters
  • Power transmitters with internal loop supply
  • Store frequently-used test setups for later use
  • Pressure measurement to 3,000 psi/200 bar using any of the 8 intrinsically safe Fluke 700PEx Pressure Modules, calibrate pressure transmitters and I/P devices
  • Pressure switch test function to capture set, reset and deadband values
  • Compact size and weight
  • Simple, push-button user interface
  • Rugged and reliable, for field use

 

Measurement Accuracy
Voltage DC
30.000 V  0.02%+ 2 counts (upper display)
10.000 V  0.02%+ 2 counts (lower display)
90.00 mV  0.02%+ 2 counts
-10.00 mV to 75.00 mV  0.025 % + 1 count (via TC connector)
Current DC
24.000 mA  0.02%+ 2 counts
Resistance
0.0 to 400.0 Ω  0.1Ω (4-wire), 0.15Ω (2- and 3-wire)
401 to 1500 Ω  0.5Ω (4-wire), 1Ω (2- and 3-wire)
1500 to 3200 Ω  1Ω (4-wire), 1.5Ω (2- and 3-wire)
Frequency
2.0 to 1000.0 CPM  0.05 % + 1 count
1.0 to 1100.0 Hz  0.05 % + 1 count
1.00 to 10.00 kHz  0.05 % + 1 count
Sensitivity  1 V peak-to-peak-minimum
Pressure
Accuracy  from 0.025% of range using any of 8 intrinsic safe pressure modules.
   (for detailed specifications refer to pressure modules in options and accessories)
   Modules available for differential, gauge, vacuum, absolute, dual and high pressure.

 

Source Accuracy
Voltage DC
100.00 mV  0.02%+2 counts
10.000 V  0.02%+2 counts
-10.00 mV to 75.00 mV  0.025 % + 1 count (via TC connector)
Current DC
24.000 mA (Source)  0.02%+ 2 counts
24.000 mA (Simulate)  0.02%+ 2 counts
Resistance
15.0 to 400.0 Ω  0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
401 to 1500 Ω  0.5 Ω(excitation current 0.05 to 0.8 mA)
1500 to 3200 Ω  1 Ω(excitation current 0.05 to 0.4 mA)
Frequency
2.0 to 1000.0 CPM  0.05%
1.0 to 1100.0 Hz  0.05%
1.00 to 10.00 kHz  0.25%
Waveform:  5 V p-p squarewave, -0.1 V offset

 

 

RTDs and Thermocouples
Measure Accuracy:
NI-120  0.2 °C
PT-100 (385)  0.33 °C
PT-100 (393)  0.3 °C
PT-100 (JIS)  0.3 °C
PT-200 (385)  0.2 °C
PT-500 (385)  0.3 °C
PT-1000 (385)  0.2 °C
Resolution:  0.1 °C
J:  0.7 °C
K:  0.8 °C
T:  0.8 °C
E:  0.7 °C
R:  1.8 °C
S:  1.5 °C
B:  1.4 °C
L:  0.7 °C
U:  0.75 °C
N:  0.9 °C
Resolution  J, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F
XK  0.6°C
BP  1.2°C
Source Accuracy:
NI-120  0.2 °C
PT-100 (385)  0.33 °C
PT-100 (393)  0.3 °C
PT-100 (JIS)  0.3 °C
PT-200 (385)  0.2 °C
PT-500 (385)  0.3 °C
PT-1000 (385)  0.2 °C
Resolution:  0.1 °C
Note:  Accuracy stated for 4-wire measurement.
J:  0.7 °C
K:  0.8 °C
T:  0.8 °C
E:  0.7 °C
R:  1.4 °C
S:  1.5 °C
B:  1.4 °C
L:  0.7 °C
U:  0.75 °C
N:  0.9 °C
Resolution  J, K, T, E, L, N, U: 0.1 °C, B, R, S: 1 °C
XK  0.6°C
BP  1.2°C

 

 

Specifications
Ramp Functions
Source functions:  Voltage, current, resistance, frequency, temperature
Ramps  Slow ramp, Fast ramp, 25% step-ramp
Loop Power Function
Voltage:  12 V
Accuracy:  10%
Maximum current:  22 mA, short circuit protected
Step Functions
Source functions:  Voltage, current, resistance, frequency, temperature
Steps  25% of range, 100% of range

 

 

Safety Specifications
Agency Approvals
   ATEX II 1 G Ex ia IIB 171°C
   I.S. Class I, Division 1 Groups

 

 

Mechanical & General Specifications
Size
   130 x 236 x 61 mm
   5.188" x 9.291" x 2.402"
Weight
   0.85 kg (1.874 lbs.)
Batteries
   4 AA alkaline batteries
Warranty
   Three year
Battery Replacement
   Separate battery compartment, accessible without breaking calibration seal
Side port connections